AirSenseTM 10 AutoSetTM
The AirSense 10 AutoSet is a premium auto-adjusting pressure therapy device that automatically adjusts pressure levels to fit your changing needs and to ensure you’re receiving the lowest pressure necessary. The AirSense 10 AutoSet is also set to deliver therapy at the most comfortable temperature and humidity levels when used with ResMed’s new HumidAir™ heated humidifier and ClimateLineAir™ heated tube.
Engineered to be intuitive and easy-to-use, you can simply plug in the AirSense 10 AutoSet device and press Start
TECHNICAL SPECIFICATIONS
90W power supply unit
AC input range: 100–240V, 50–60Hz 1.0–1.5A, Class II
115V, 400Hz 1.5A, Class II (nominal for aircraft use)
DC output: 24V 3.75A
Typical power consumption: 53W (57VA)
Peak power consumption: 104W (108VA)
Sensors
Pressure sensor:
Internally located at device outlet, analog gauge pressure type, -5 to +45 cm H2O
Flow sensor:
Internally located at device inlet, digital mass flow type, -70 to +180 L/min
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H2O for more than 6 sec or 40 cm H2O for more than 1 sec.
Sound
Pressure level measured according to ISO 17510-1:2007 (CPAP mode):
SlimLine:
26.6 dBA with uncertainty of 2 dBA
Standard:
Physical - device and water tub
Dimensions (H x W x D):
4.57" x 10.04" x 5.91" (116 mm x 255 mm x 150 mm)
Air outlet (complies with ISO 5356-1:2004): 22 mm
Weight (device and standard water tub):
44 oz (1248 g)
Temperature
Maximum heater plate:
154°F (68°C)
Cut-out:
165°F (74°C)
Maximum gas temperature:
106°F ( 41°C)
Air filter
Standard:
Material: Polyester non woven fiber
Average arrestance: >75% for ~7 micron dust
Hypoallergenic:
Material: Acrylic and polypropylene fibers in a polypropylene carrier Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel.
Wireless module
Technology used:
CDMA (USA and Canada only) 2G GSM (all regions except USA and Canada)
FCC ID: 2ACHL-AIR10CD
The AirSense 10 device complies with FCC Rules.
The AirSense 10 device should be used at a minimum distance of 0.8" (2 cm) from the body during operation. Additional information regarding the FCC Rules for this device can be found on www.resmed.com.
Operating pressure range
AutoSet, AutoSet For Her, Elite, CPAP 4 to 20 cm H2O
Supplemental oxygen
Maximum flow: 4 L/min
Pneumatic flow path
We provide CPAP, BIPAP, BPAP, VPAP, CPAP Mask, BIPAP Mask, Humidifier, Oxygen Concentrator, Ventilator, Defibrillator, ECG Machine and Wheelchair at lowest price in India. Rent Facility is also available for CPAP, BIPAP and Oxygen Concentrator.